Uso De Valvula

Páginas: 3 (514 palabras) Publicado: 13 de agosto de 2012
AN74i VENT PUMP FAILURE DOES NOT POSE ETHYLENE OXIDE EXPOSURE RISK
This study1 was conducted to determine the consequences of a ventilation pump failure in an Andersen AN74i sterilizer. Ofparticular concern was possible operator exposure to ethylene oxide (EO) in the vicinity of the sterilizer. Measurements taken at the cycle’s peak concentration of EO resulted in an undetectable level (< 1part per million). CONDITIONS: The temperature inside the sterilization liner bag measured at 24.3 to 25oC. The relative humidity ranged from 42.4 to 60.5 percent. The test room itself had no airchanges. Performing the test in an unventilated room increases the risks associated with a malfunction of this kind. Andersen Products, Inc. recommends that the sterilizer be installed in a room with atleast 10 air changes per hour. Operation of the sterilizer in a room with no air changes constitutes a gross misuse of the system. MATERIALS:
AN74i Sterilizer PAN-TY cable ties and Thomas & Betts cabletie tool Andersen Sniffer™ with nylon gas sampling bags Gas sampling pump and collection bag Shimadzu gas chromatograph and analyzer 1.0ml gas-tight syringes for gas injection into GC Standard Load: •10 AN10 Andersen Tubes sealed in 4.5” PolyEthylene/PolySurlyn pouch • 2 Patient Gowns wrapped in CSR wrap • 1 AN42 Sump Pump wrapped in CSR wrap • 6 Pairs of Latex Gloves sealed in Seal and Peel •10 Cotton-Tipped Applicators sealed in Seal and Peel • 30 PPE Sutures inserted in aluminum pouches sealed in a self-seal 7”x13” paper plastic pouch • 4 Hemostats sealed in Seal and Peel • 12 Syringes(3 large, 3 medium and 6 small) sealed in a self-seal 7”x13” paper plastic pouch • 10 Glass Vials (amber with rubber stoppers) sealed in a self-seal 7”x13” paper plastic pouch • 1 x Humidichip placedin a Humiditube • Cox Recorder • AN79 (17.6ml) Anprolene Ampoule

occurred inside the liner bag at 2.5 hours after activating the ampoule. This test simulated a vent pump failure 2.5 hours into...
Leer documento completo

Regístrate para leer el documento completo.

Estos documentos también te pueden resultar útiles

  • Uso de válvulas reductoras de presion
  • Valvulas
  • Valvulas
  • valvulas
  • Valvulas
  • Valvulas
  • valvulas
  • Valvulas

Conviértase en miembro formal de Buenas Tareas

INSCRÍBETE - ES GRATIS