DI-160 Service Manual
Edition 3
rd

2. Setup & Maintenance
2.1. Initial Set-up Procedure 1. Remove the rear lid removing 4 screws. 2. Remove a shield sheet, which covers over the PCB. 3. Solder load cell wires to specified points of PCB. Silk on PCB Signals CNW-6410 BLACK Green/Yellow Green Yellow Red Orange Blue White Fig. Locations on Main PCB

BLK -------------- F.G. *SLD-----------*GRN ----------*YEL -----------*RED-----------RS- -------------F.G. Load cell Output (+) Load cell Output Load cell Applied Voltage (+) Sensor (-)

RS+ -------------- Sensor (+) *WHT --------GND ---------Load cell Applied Voltage (-) GND for Mercury SW

HGSW ---------- Input of Mercury SW NOTES: When connecting DIGI load cell, wires * marked should be soldered at each soldering point as shownright. 4. Insert powers cord from AC adapter into Power Jack on PCB. 5. Pass the load cell wire and power cord through the path in rubber shield block

6. Put the shield sheet and the lid in its position
2.2. Start-up Next, specification data and weight calibration are required to get the indicator working properly. 1. Press [ON/OFF] key to start up the machine. Software version is displayed suchas “Ver.0.04”. All segments like “8, 8, 8, 8, 8, 8” are displayed.

The display blanks off. The display locks up with all “8”s. Now it is ready to set up the specification data and weight calibration. * If “Lo-VoL“ is displayed, this indication that the input power voltage is too low to operate the unit. This is because the input voltage drop detection circuit is equipped to prevent the unitfrom malfunctioning when the voltage dropped. To exit this error, please reset the unit by turning off and on.

3

DI-160 Service Manual
Edition 3
rd

2.3. SPEC’s Data setting for RS-232C Communication 1. Press [ T ] key 3 times by holding [REZERO] key --------- “Z” + “TTT” This key operation leads you to SPEC data mode regarding RS-232C Communication. 2. For setting procedures, please referto the following pages. 3. To exit this mode, press [ T ] key . 2.4. SPEC’ s Data Setting for Weigh & Measure 1. Turn on the SPAN SW. (SW6 on Main Board.) 2. Press [ ] key 3 times by holding ” SPAN SW A/D Block TPB-2575

[Re-Zero] ------ “ Z “ + “

This key operation leads you to SPEC setting mode regarding W&M. 3. For setting procedures, please refer to the following pages. 4. To exit thismode, press [ T ] key. 2.5. How to Set /Change SPEC Data As introduced at the previous section, there are 2 kinds of key-operation when entering SPEC setting mode. a . [Re-Zero] + [ T ] [ T ] [ T ] b . [Re-Zero] + [ ][ ][ ------- for SPEC No. 0-No. 3

] -------for SPEC No. 4-No. 15

In case of “b” operation , the following display will appear.

Bit3

Bit2

Bit1

Bit0

04
SPEC No.1

1 0

0

SPEC data

*The two leftmost bits indicate SPEC No. such as “04 “ *The remaining 4 bits indicate SPEC numeric data such as “ 1100 “. *Mark “ “ under a figure indicates the current programmable position

When setting or changing SPEC data, key switches function are as follows. [REZERO] [T] [ [ ] ] ----------SPEC No. skips each time this key is pressed.

------------ To renewthe data and exit this mode. ------------ To move digit position. “ ”will move to next digit. ------------ To change SPEC data to either “0” or “1”.

From the next page, the SPEC data list is attached so that you can set/change the data according to your requirement.

4

DI-160 Service Manual
Edition 3
rd

2.6. SPEC Data List a. [Re-Zero] + [ T ] [ T ] [ T ] SPEC Bit 3 RS-232CConnection SPEC 00 0: Not Connected 1: Connected Stop Bit SPEC 01 0: 1 bit 1: 2 bit Bit 2 Baud Rate [bps] 000: 1200 001: 2400 010: 4800 Data Length 0: 7 bits 1: 8 bits 011: 9600 000: 19800 011~111: Not Used Parity Bit 0 0: No Parity 0 1: Odd Stable of Text Data 0: Inhibit 1: Allow 1 0: Even 11: Not Used Header of Text Data 0: Without 1: With Bit 1 Bit 0

Selection of RS-232 Communication Type SPEC 02...
Leer documento completo

Regístrate para leer el documento completo.

Estos documentos también te pueden resultar útiles

  • ds 160
  • DS 160 tutorial
  • Tutorial ds-160
  • Ds 160
  • DS 160
  • Online Nonimmigrant Visa Application (Ds- 160)
  • DS 160 powerpoint spanish PARA VISAS
  • formulario ds 160

Conviértase en miembro formal de Buenas Tareas

INSCRÍBETE - ES GRATIS