Handy Calibrators
S ource and measure operations can be performed at the same time.
(Select from the following source signal and measurement signal options: voltage,
current, resistance, thermocouple (TC), resistance temperature detector (RTD),
frequency, pulse).
A C voltages, including supply voltage, can be measured.
I ncludes a wide array of additional functions.
E asy operation.
Compact size and Lightweight

Actual Size

Bulletin CA71E

Handy

Calibrators

Comprehensive source/measure instruments for all your calibration and
equipment checking needs at maintenance sites
Sensors and instruments require a wide variety of operation checks in the field. The CA71
Handy Calibrator is a small, lightweight, multifunction calibrator that can simultaneously source
andmeasure voltage, current, resistance, TC, RTD, frequency, and pulse signals. The CA51
Handy Calibrator is a basic model with the same functions as the CA71 except TC and RTD
measurement functions and online communication functions.

Simultaneous signal source and
measurement capability
The CA71 lets you handle regular tests on TCs, RTDs and various
other types of sensors and instruments, aswell as operation
checks when a problem has occurred. By itself it can source
signals for input to equipment, and check output signal from
equipment. With the CA71, you don't need to worry about
synchronizing operations with other setting devices.
(TC and RTD measurement functions are CA71 only.)

A wide array of useful functions
Source
Values set in steps of 4-20 mA
24V DC Power Supplyto Transmitter
Divided output (n/m) function
Output settings are divided, eliminating the need for bothersome
calculations for percentage output.
Autostep function
Changes the output value in step form based on the setting from
the divided output (n/m) function. Changes can be sourced
automatically every 10% or 25%.
Online communication (CA71 only)
RS-232C-compliant optically isolatedinterface
Sweep function
Linearly increases or decrease the output. The
increasing/decreasing time can be set to either 16 or 32 seconds.
Memory function
Source values and measurements forming individual value sets
can be saved to or read from the Handy Calibrator's internal
memory (maximum 50 value sets).
Temperature monitor function

AC voltage (including supply voltage)
measurementcapability

Easy operation

In cases where numerous signal converters and other devices are
mounted on a rack or panel, the Handy Calibrator can be used to
check the input and output signals of each device, while
simultaneously checking the power supply. There is no need for a
separate multimeter to measure supply voltage.

The Handy Calibrator incorporates rotary switches for simplehandling. Just open the carrying case cover and connect the
cables, and you're ready to take measurements.

Panel layout
2. Terminal adapter
99021 (*2)

1. Measurement range
setting rotary switch
Select from the following:

Measurement input terminals

4. Signal generation range
setting rotary switch
Select from the following:

*2: Not used on CA51.

3. Output terminals for
signalgeneration

bAC voltage

bDC voltage

bDC voltage

bTC

bTC (*1)

bResistance

bResistance

bRTD

bRTD

bFrequency
pulse

(*1)

bFrequency
pulse

bDC current

bDC current
*1: Only for CA71

5. POWER key

bUp/Down keys for generated signal setting
bDivided output (n/m) function
bMemory function

6. Various function settings

7. LIGHT key

8. TEMP keyTurns on the LCD
backlight.

Enables temperature
monitoring.

bSpare parts
Product

Model

Source signal lead cable

Measurement lead cable

98020

Carrying case

RD031

Remarks One set, including one red
cable and two black cables.
Length: approximately 1.7
meters

One set, including one red
cable and one black cable.
Length: approximately 1.0
meter

Terminal adapter...
Leer documento completo

Regístrate para leer el documento completo.

Estos documentos también te pueden resultar útiles

  • INTOUCHABLES
  • Intouchable
  • Análisis "Intouchables"
  • intouch
  • Intouch
  • Intouch
  • labview vs Intouch
  • Comunicación tsx micro con intouch

Conviértase en miembro formal de Buenas Tareas

INSCRÍBETE - ES GRATIS