Despirogenización Tunnel
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38 Depyro-Tunnel
Vessel: Product: Time Zone:
Tunel K1 Ampul 20ml Środkowoeuropejski czas stand. (UTC offset 01:00:00)
Comments
Name: Comment Related to: Comment:
Introduction Depyro-Tunnel
Depyrogenation tunnel K1
Made by loggers and sensorsTrackSense Pro from Ellab A / S
DEMO File from ELLAB webpage.
Operator/Date:
- Generated by ValSuite Pro ver. 2.6.25.0 Printed: 2010-02-23 14:36:06 Page 1 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session Text:
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38 Depyro-Tunnel
Vessel: Product: Time Zone:
Tunel K1 Ampul 20ml Środkowoeuropejskiczas stand. (UTC offset 01:00:00)
Graph
Operator/Date:
- Generated by ValSuite Pro ver. 2.6.25.0 Printed: 2010-02-23 14:36:06 Page 2 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session Text:
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38 Depyro-Tunnel
Vessel: Product: Time Zone:
Tunel K1 Ampul 20ml Środkowoeuropejski czas stand. (UTC offset01:00:00)
Time Markers
Time Stamp: 00:25:48 00:40:54
Name: Start of the test End Test
Description: Placement of the tunnel Removal of the tunnel
Operator/Date:
- Generated by ValSuite Pro ver. 2.6.25.0 Printed: 2010-02-23 14:36:06 Page 3 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session Text:
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38Depyro-Tunnel
Vessel: Product: Time Zone:
Tunel K1 Ampul 20ml Środkowoeuropejski czas stand. (UTC offset 01:00:00)
Channel Overview
Sensor ID 10188-1 10188-2
Name
Logger ID
E-Val Flex S/N N/A N/A
Description Sensor placed in an ampoule Sensor placed in an ampoule
Sample Rate 1 (hh:mm:ss) 00:00:01 00:00:01
Sample Rate 2 (hh:mm:ss) N/A N/A
The left side of the tunn12783 The right side of the tun 12783
Operator/Date:
- Generated by ValSuite Pro ver. 2.6.25.0 Printed: 2010-02-23 14:36:06 Page 4 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session Text:
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38 Depyro-Tunnel
Vessel: Product: Time Zone:
Tunel K1 Ampul 20ml Środkowoeuropejski czas stand. (UTC offset01:00:00)
Unit Configuration:
Position 1: None Lethality Auto Start/Stop Operator/Date:
Position 2: N/A N/A
None
- Generated by ValSuite Pro ver. 2.6.25.0 Printed: 2010-02-23 14:36:06 Page 5 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session Text:
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38 Depyro-Tunnel
Vessel: Product: Time Zone:Tunel K1 Ampul 20ml Środkowoeuropejski czas stand. (UTC offset 01:00:00)
Name: Description: Position: Sensor S/N: 1 2 None None
Tunel Ampułki 20 ml Position Name: Tunel left Tunel right
Type:
Retort
Position Description: Sensor in an ampule Sensor in an ampule
Sensor Alias:
Sample Rate: 00:00:01 00:00:01
Operator/Date:
- Generated by ValSuite Pro ver. 2.6.25.0 Printed:2010-02-23 14:36:06 Page 6 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session Text:
DEMO Depyrogenation 2008-03-27 09:54:03 2008-03-27 10:39:38 Depyro-Tunnel
Vessel: Product: Time Zone:
Tunel K1 Ampul 20ml Środkowoeuropejski czas stand. (UTC offset 01:00:00)
Lethality Overview
Ref. Temperature: Min. Temperature : Start Date : Name: Lethality - The left sideof the tunn Lethality - The right side of the tun
250,00 °C 250,00 °C 2008-03-27 Description:
Z-Value: D-Value:
46,4 °C 5
Start Time: 00:25:48 00:25:48
End Time: 00:40:54 00:40:54
Max. Value: 48,66 65,79
Operator/Date:
- Generated by ValSuite Pro ver. 2.6.25.0 Printed: 2010-02-23 14:36:06 Page 7 of 19
Operator: Process: Session Start: Session Stop: Session Name: Session...