Calibracion bascula dwm-iv
FSE F6
F6.0 Weight Units used for Printing d - No weight units printed.
0 - Lb
1 - kg
2 - g
3 - Ton
4 - Tonne (metric ton)
5 - oz
6 - oz troy
F6.1Increment Size
Least significant digit increment size. Choices: 1, 2, 5, 10, 20, 50, 100
F6.2 Decimal Position
Display decimal position from none to four places to the left.
0 - No decimal1 - .X
2 - .XX
3 - .XXX
4 - .XXXX
F6.3 Enter CAL Mode
d - Continue to last step (CAL d) E - Enter CAL mode.
F6.4 Linearity Factor
Corrects non-linearity in weigh system. Thevalue required is determined empirically. Amount of correction depends upon perce nt usage of the load cell as well as the system capacity and resolution.
Choices: 7 to -7
E SCALE Empty ScaleRemove all weight from the scale and then press the [PRINT] button.
888888 Full Scale Capacity
888888 indicates the previously entered Full Scale Capacity in even units. Use the numerickeyboard to enter the Full Scale capacity. Press the [ENTER] button to advance to the next step.
15 SEC Zero Setting Fifteen Second Delay
Counts down from 15 to zero to capture the zero setting. Ifthe [SELECT] button is pressed during the time-out period, the meter will immediately terminate the delay and ignore the motion criterion to capture the weight. Allow at least three (3) secondsafter weight has been applied to the scale before pressing [SELECT] or the weight captured may be erroneous.
Add Ld Add Load or Add ALL
The weight applied must be at least 5% of the full scalecapacity. The message Add ALL will appear if Linearity Factor is not zero indicating that Full Scale Capacity must be applied.
[PRINT] - Advances to next step if span must be set.
[TARE] -Establishes a new zero calibration setting without changing the span calibration previously set. The display will show CAL d. This feature is disabled if the Add ALL message is displayed.
888888 Tes t...