Estudio de tolerancia en micropipetas

Páginas: 2 (321 palabras) Publicado: 24 de mayo de 2011
Estudio de medidas de desviación en Micropipetas
Micropipeta: 10-100 uL
Medida | Rainin | Axygen | Eppendorf | Global | Belfast |Media |
10 | 0.35 / 3.5 | 0.16/1.6 | 0.3/3 | 0.18/1.8 | | 0.25 |
50 | 0.4/ 0.8 | 0.4/0.8 | 0.5/1 | - | 0.6/1.2 | 0.48 |
100 | 0.8/ 0.8| 0.8/0.8 | 0.8/0.8 | 0.6/0.6 | 1/1 | 0.8 |

Micropipeta: 20-200 uL
Medida | Rainin | Axygen | Eppendorf | Global | Belfast | Media |20 | 0.5/2.5 | 0.24/1.2 | 0.5/2.5 | | 0.3/1.5 | 0.385 |
100 | 0.8/0.8 | 0.8/0.8 | 1/1 | | 1.0/1.0 | 0.9 |
200 | 1.6/0.8 | 1.2/0.6| 1.2/0.6 | | 2.0/1 | 1.33 |

Micropipeta: 100-1000 uL (ul/%)
Medida | Rainin | Axygen | Eppendorf | Global | Belfast | Media uL |100 | 3.0/3.0 | 1.6/1.6 | 3/3 | 1.6/1.6 | 1.0/1.0 | 2.05 |
500 | 4.0/0.8 | 3.5/0.7 | 5/1 | | 5.0/1.0 | 4.375 |
1000 | 8.0/0.8 |6/0.6 | 6/0.6 | 3/0.3 | 10.0/1.0 | 6.6 |

Micropipeta: 500-5000 uL
Medida | Rainin | Axygen | Eppendorf Research | Global | Belfast |Media |
500 | 12/2.4 | 6/0.6 | 12/2.4 | | 5.0/1.0 | 8.75 |
2500 | 15/0.6 | 15/0.6 | 30/1.2 | | | 20 |
5000 | 30/0.6 | 25/0.5 | 30/0.6| | 50/1 | 28.3 |

Micropipeta: 1 a 10 mL
Medida | Rainin | Axygen | Eppendorf R | Global | Belfast | Media |
1 | 50/5.0 | 25/2.5| 30/3 | | 10.0/1.0 | 28.75 |
5 | 50/1.0 | 40/0.8 | 40/0.8 | | 50/1 | 45 |
10 | 60/0.6 | 50/0.5 | 60/0.6 | | 80/0.8 | 62.5 |
Leer documento completo

Regístrate para leer el documento completo.

Estos documentos también te pueden resultar útiles

  • Micropipetas
  • micropipetas
  • Micropipeta
  • Que Es Una Micropipeta
  • Micropipetas
  • Micropipetas
  • Micropipetas
  • Micropipetas

Conviértase en miembro formal de Buenas Tareas

INSCRÍBETE - ES GRATIS