Preferred Room Layout
The layout shown below is based upon a typical equipment configuration and should be considered as a general design guideline. Site conditions, applicationrequirements, customer preferences, and/or equipment configuration may significantly impact suite design and equipment layout. It is recommended to request site-specific drawings from a Philipsrepresentative early in the design process.
6'-11" [2108mm]
18'-11" [5766mm] 6'-4" [1930mm]
ISO 6'-6" [1981mm] Gantry & Patient Table Longitudinal UPS Isocenter WT WS
COM
Optional deskavailable from Philips
Equipment Layout
Recommended Ceiling Height: 8'-0" (2438mm) Minimum Ceiling Height: 7'-6" (2286mm)
0 0 1' 2' .5m 1m 4' 2m 8' 12' 4m
Tabletop available from Philips
EW Finallocation to be determined. Coordinate with local Philips Service.
14'-0" [4268mm]
OC
Equipment Legend A Furnished and installed by Philips B Furnished by customer/contractor and installedby customer/contractor C Installed by customer/contractor D Furnished by Philips and installed by contractor E Existing F Future G Optional item furnished by Philips Equipment Designation DescriptionWeight lbs [kg] 4078 [1850] 756 [343] 25 [11] 331 [150] 604 [274] 123 [56] 90 [41] Heat Load Btu/hr [W]
A A A
WS WT OC
MX 16-Slice CT Scanner Gantry Patient Table Operators Console
17075[5004] 2561 [751] 2210 [648] 375 [110] 1000 [293]
A COM Computer Rack (under desk) G G G
ISO UPS EW
LM Isolation Transformer Power Unit Console UPS Extended Brilliance WorkspaceEnvironmental Requirements for General Equipment Locations
The above conditions must be maintained at all times including overnight, weekends, and holidays. Heat output in one area of the CT Suite must notaffect temperature and humidity in other areas. It is strongly recommended that any definable areas within the suite, i.e. equipment closets, control areas, etc. (if applicable), be individually...